Defibrillator |
|
Defibrillator Model |
HeartStart M5066A |
|
Defibrillator Family
|
HSI |
| How
Supplied
|
Defibrillator,
Owner's Manual, battery, 1 adult SMART
Pads cartridge, Quick Reference Guide,
Quick Start poster, and your choice of
standard or slim carry case. |
|
Waveform |
Truncated Exponential
Biphasic. Waveform parameters adjusted
as a function of patient impedance. |
|
Energy
|
Single energy output.
Adults: nominal 150 Joules into a 50 ohm
load.
Infant/Child: nominal 50 Joules into a
50 ohm load.
Automatically set based on type of SMART
Pads cartridge installed. |
|
Shock-to-Shock Cycle Time
|
Typically less than
20 seconds between shocks in a series. |
| Quick
Shock |
Able to deliver a
shock after the CPR pause typically in
eight seconds. |
| Voice
Instructions
|
Detailed voice
messages guide responder through use of
the defibrillator. |
| CPR
Voice Coaching
|
Voice instructions in
adult and infant/child CPR available at
user's option. |
| Shock
Delivery
|
Via adhesive pads
placed on patient's bare skin as
illustrated on pads. |
|
Controls
|
Green SMART Pads
cartridge handle, Green On/Off button,
Blue i-button, Orange shock button. |
|
Indicators
|
Ready light; Blue i-button;
Caution light. |
Size
Height
Width
Depth
|
2.8 inches (7 cm)
8.3 inches (21 cm)
7.4 inches (19cm)
|
Weight
With battery and pads cartridge
Without battery or pads cartridge
|
3.3 lbs (1.5 kg)
2.4 lbs (1 kg) |
|
Sealing
|
Drip proof per
EN60529 class IPX1.
Solid Objects per EN60529 class IP2X. |
|
Temperature
|
Operating: 32º - 122º F (0º - 50º C)
Standby: 50º - 109º F (10º - 43º C) |
|
Humidity |
Operating: 0% to 95% relative,
non-condensing
Standby: 0% to 75% relative,
non-condensing |
|
Altitude
|
Operating: 0 to 15,000 feet
Standby: 0 to 8,500 feet >48 hours
and 8,500 to 15,000 feet <48 hours |
|
Shock/Drop Abuse
|
Withstands 1 meter drop to any edge,
corner or surface. |
|
Vibration
|
Meets
EN1789 random and swept sine, road
ambulance specification in operating and
standby states. |
| EMI
(Radiated/Immunity)
|
Meets
EN55011 Group 1 Level B Class B and
EN61000-4-3. |
|
Patient Analysis
|
Evaluates
patient ECG and signal quality to
determine if a rhythm is shockable and
evaluates connection impedance for
proper defibrillation pad contact.
Rhythms considered shockable are
ventricular fibrillation (VF), and
certain ventricular tachycardias (VT)
associated with lack of circulation,
including ventricular flutter and
polymorphic ventricular tachycardia. For
safety reasons, some VT rhythms
associated with circulation will not be
interpreted as shockable, and some very
low-amplitude or low-frequency rhythms
will not be interpreted as shockable VF. |
|
Sensitivity/Specificity
|
Meets
AAMI DF80 guidelines and AHA
recommendations for adult
defibrillation(Circulation
1997;95:1677-1682). |
|
Artifact Detection
|
The
effects of pacemaker artifact and
electrical noise are minimized with
artifact detection. |
| Type
|
9 Volt
DC, 4.2 Ah, composed of disposable
long-life lithium manganese dioxide
primary cells. |
|
Capacity
|
Minimum 200 shocks or 4 hours of
operating time. |
|
Install-By Date
|
Battery
is labeled with an install-by date of at
least five years from date of
manufacture. |
|
Standby Life
|
Four
years typical when battery is installed
by the install-by date.(Will power the
AED in standby state within the
specified standby temperature range,
assuming one battery insertion test and
no defibrillation uses.) |
|
Adult
SMART Pads
Cartridge |
M5071A defibrillation
pads for patients 8 years of age and
older or 55 lbs. (25 kg) and over. |
|
Infant/Child SMART Pads Cartridge |
M5072A defibrillation
pads for patients under 8 years of age
or 55 lbs. (25 kg). |
|
Energy
Delivered |
Adult: nominal 150
Joules into a 50 ohm load.
Infant/Child: nominal 50 Joules into a
50 ohm load. |
|
How
Supplied
|
Disposable cartridge,
containing adhesive defibrillation pads,
clicks into defibrillator for an
integrated pads solution. |
|
Active
Surface Area |
13.2 in2
(85 cm2 ) each |
|
Cable
Length |
Adult
pads: 54 inch (137.1 cm)
Infant/Child pads: 40 inch (101.6 cm) |
|
Use-By
Date |
Cartridge
is labeled with a use-by date of at
least two years from date of
manufacture. |
|
Adult Training Pads Cartridge |
M5073A
|
|
Infant/Child Training Pads Cartridge |
M5074A |
|
Function
|
Special
pads place HeartStart OnSite into
training mode and disable its energy
delivery capability. Feature 8
real-world training scripts. |
| Daily
Automatic Self-tests
|
Tests internal
circuitry, waveform delivery system,
pads cartridge and battery capacity. |
| Pads
Integrity Test
|
Specifically tests
rediness- for use of pads (gel
moisture). |
|
Battery Insertion Test
|
Upon battery
insertion, extensive automatic
self-tests and user-interactive test
check device readiness. |
|
Status Indicator
|
Blinking green Ready
light indicates ready for use. |
|
Infrared
|
Wireless transmission
of event data to personal computer using
the IrDA protocol. |
|
HeartStart Event Review Software
|
Data management
software (optional) for download and
review of data retrieved through
defibillator's infrared data port. |
| Data
Stored
|
First 15 minutes of
ECG and the entire incident's events and
analysis decisions. |